在线联系治疗仪网—中频治疗仪,脉冲磁治疗仪,紫外线治疗仪,骨质增生治疗仪

 

  中频治疗仪药物离子导入机

中频治疗仪 | 中频治疗仪(家用型) | 家用型中频治疗仪 | 中频治疗仪(漫游温热)型 | 漫游温热中频治疗仪 |中频治疗仪(中频+直流)型 | 中频+直流中频治疗仪 | 中频治疗仪(直流型) | 直流中频治疗仪 | 中频治疗仪(漫游温热)型 | 漫游温热中频治疗仪

   

    

  

中频治疗仪药物离子导入机

1、电源电压:220V 50HZ 功率:15W
2
、内存处方:20 调幅度:100%100
3
、输出方式:A:电脑中频(双向)B:药物导入(单向)C:加热B常温(不加热)
4
、输出波形:矩形波、尖波、三角波、指数波、棱形波、阶梯波、正弦波、锯齿波(前、后)、积分波、微分波等。

1、性能结构

治疗仪由CPU电路板、功率放大器、输出保护电路、稳压电源、治疗电极、输出线及输出线插头组成。

Therapeutic apparatus by CPU circuit board, power amplifier, the output protection circuits, regulated power supply, treatment electrode, the output line and output line component plugs.

 

该产品采用工作频率为2000-8000Hz、调制频率为0-200Hz,波形选用正弦波、方波、三角波、指数波等,由单片机编程,以等幅、调制、条幅、单波形和多波形组合输出治疗方式,除中频电疗一般生物物理特征外,增强了电刺激生物学动力效应。

The product operates at 2000-8000Hz, modulation frequency of 0-200Hz, sine wave waveform selection, square, triangle wave, the index wave and so on, by the single-chip programming to amplitude modulation, banner, a single waveform and many combination of the output waveform of treatment, with the exception of frequency electrotherapy general bio-physical characteristics, electrical stimulation enhances the biological effects of power.

其主要生物学治疗学作用为:

Its main biology therapeutics function is:

镇痛、消炎、消肿、促进局部血液循环、锻炼神经肌肉、调整神经系统功能等作用。

Analgesic, anti-inflammatory, reduces swelling, to promote local blood circulation, neuromuscular training, adjust the role of the nervous system functions.

2、技术指标    

1、工作频率:2000Hz-8000Hz频率允许误差±10%

Operating frequency: 2000Hz-8000Hz frequency of permissible error ± 10%

输出波形:(低频调制后)方波、正弦波、三角波、指数波、等幅波

Output wave shape: (after low frequency modulation), square-wave, sine wave, triangular wave, exponential wave, persistent waves

脉冲宽度:(0.25ms0.0625ms)±10%

Pulse width: (0.25ms ~ 0.0625ms)±10

最大幅度:(0100%)±10%

Largest scale: (0~100%) ±10%

负载阻抗的变化对上述这些参数无影响。

Changes in load impedance of the above had no effect on these parameters.

2、输出电流:在基准负载下500Ω±5℅,最大输出电流90mA±10(连续可调)

Output current: At baseline load 500Ω ± 5 ℅, maximum output electric current 90mA±10℅ (continuously adjustable)

3、调制频率范围:(0200Hz

Modulation frequency scope: (0~200) Hz

1、  短路承受能力:在输出最大时电极短路5s,开路5s,关闭输出1min,设备能正常工作。

The bearing capacity of short-circuit: the output at the maximum short-circuit when the electrode 5s, open-5s, turn off output 1min, equipment can work properly.

2、  输出电极温度:不大于50℃(敞开测试)

the output electrode temperature: not more than 50 (open test)

6、使用环境温度:5℃~40   湿度:≤80

The use of ambient temperature: 5 ~ 40 Humidity: ≤ 80 ℅

7、电源电压:AC220V±22V      50Hz±1Hz  

Supply voltage: AC220V ± 22V 50Hz ± 1Hz

8、整机功率:单路机≤40 VA      双路机≤80VA

Machine power: single-road machine ≤ 40 VA dual machine ≤ 80VA

9、主机尺寸:单路300×220×90mm        双路400×320×160mm

Main engine size:: single 300 × 220 × 90 (mm) dual 400 × 320 × 160 (mm)

10、安全类型:1BF

Security Type: Class 1 BF type

11、熔断器规格:单路F5×20B×1.5A2只)

                双路F5×20B× 3A2只)

Fuse specifications: single road F 5 × 20B × 1.5A (2 rats)
                      Dual F
5 × 20B × 3A (2 rats)

 


廊坊市治疗仪有限公司

地址:河北省廊坊市建设北路63号 电话:0316-2229110 传真:0316-2229110 信箱:5122398@163.com  网址:http://www.263118.cn/

本页关键词:治疗仪 紫外线治疗仪 中频治疗仪 脉冲磁治疗仪

 



地址:河北廊坊市建设北路63号  冀ICP备09013213号
电话:0316-2229110 15803168811 传真:0316-2229110 邮箱:5122398@163.com  
版权所有 紫外线治疗仪 中频治疗仪 脉冲磁治疗仪 旋磁按摩仪—廊坊治疗仪有限公司